Water cooling option

Cooling chiller/Nitrogen gas generator

In Peltier temperature testing products with water cooling methods, options such as the following may be required depending on the customer’s environment.

Cooling chiller/ CCA-1112
冷却チラー CCA-1112
冷却チラーCCA-1112図面

Feature

  • The temperature adjustment system and cooling capacity are based on the rated power supply voltage, 50Hz, room temperature of 20℃, and no-load values
  • The cooling capacity is 土5% of the displayed capacity (model CCA-1112).

Specification

Type CCA-1112
Circulation method Closed system circulation
Temperature control range (℃) -20 ~ +20
(Without heater)
Temperature control accuracy (℃) ±2
Refrigeration equipment and refrigerants 350W R404A
Cooling capacity (W): 10℃ 450 (385kcal/h)
Cooling capacity (W) :-10℃  310 (266kcal/h)
External Circulation Capability Max. flow rate 14 / 14.5 (L/min)
Max. lifting height 4.2 /5.6m(50/60Hz)
External dimension(mm) 232(W) x 497(D) x 490(548)(H)(Nozzle tip)
Weight (kg) 29
Power Input/ Power Supply 8A 800VA AC100V 50/60Hz

 

Nitrogen gas generator TWC-05
TWC-05
窒素ガス発生装置TWC-05図面

Feature

  • Due to the built-in compressor, nitrogen gas is generated only with AC100V.
  • The compressor is of the oil-free type, so no maintenance such as oiling is required.
  • Air filter and mist filter are standard equipment.
  • Nitrogen gas concentration adjustment range is 90 to99.9%.

Specification

Type TWC-05
Generate method Membrane separation
Environment temperature (℃) 5 ~ 45
Environmental humidity Lower than 85%RH (Indoor)
Production amount (NL/min) 0.5 (Conditional.)
Nitrogen concentration (%) 99.9 (Conditional.)
Max. Power consumption (W) 200
External dimension (mm) 400(W) x×370(D) x 360(H)
(Excluding protrusions)
Send pressure (MPa) 0.6
(Internal compressor)
Weight (kg) 20
Power Input / Power Supply AC 100
Anti-condensation cover
結露防止カバーA-CO2
    • Dedicated anti-condensation cover for air/water-cooled plates.
    • The cover is just the right size to block outside air, and the nitrogen induced from the nitrogen outlet prevents condensation on the test object.
    • Clear type allows you to visually see changes in the test object.

Product Catalog

For any questions regarding the product or product design tailored to specifications, please feel free to contact us.

Click here to download the catalog.